Remote Control Commands
R&S
®
SMB100A
450Operating Manual 1407.0806.32 ─ 21
Table 7-2: Cross-reference between the manual and remote control
R&S name SCPI name Command under manual con-
trol
AUTO IMMediate "Auto" mode
SINGle BUS "Single" mode.
EXTernal EXTernal "Ext Single" and "Ext Step" mode.
Use command LFO:SWEep:MODE
to select between the two sweep
modes.
EAUTo - "Ext Start/Stop" mode.
:TRIGger<hw>:FSWeep:SOURce....................................................................................450
:TRIGger<hw>:FSWeep[:IMMediate]................................................................................451
:TRIGger<hw>:LFFSweep.............................................................................................. 452
:TRIGger<hw>:LFFSweep:SOURce.................................................................................452
:TRIGger<hw>:LFFSweep:IMMediate.............................................................................. 452
:TRIGger<hw>:PSWeep:SOURce....................................................................................453
:TRIGger<hw>:PSWeep[:IMMediate]................................................................................453
:TRIGger<hw>[:SWEep]:SOURce....................................................................................454
:TRIGger<hw>[:SWEep][:IMMediate]................................................................................455
:TRIGger<hw>[:IMMediate]............................................................................................. 455
:TRIGger<hw>:FSWeep:SOURce <Source>
Sets the trigger source for the RF frequency sweep.
The names of the parameters correspond directly to the various settings under manual
control. SCPI uses other names for the parameters; these names are also accepted by
the instrument. The SCPI names should be used if compatibility is an important consid-
eration.
An overview of the various names is given in Table 7-2.
TRIGger Subsystem