Instrument Function
R&S
®
SMB100A
111Operating Manual 1407.0806.32 ─ 21
â–º To access this dialog, press the SETUP or MENU key and select "Remote > Instru-
ment Emulations".
The "Instrument Emulations" dialog enables you to emulate a remote control com-
mand set of several other signal generators.
The remote commands required to remotely configure the emulation settings are
described in Chapter 7.15, "SYSTem Subsystem", on page 433.
Language
Selects the instrument whose remote command set is emulated by the R&S SMB.
Remote command:
:SYSTem:LANGuage on page 443
Mode
Selects the way the instrument identification is performed.
"Automatic"
Sets the "IDN String" and the "OPT String" automatically for the
instrument selected with the parameter Language.
"User Defined"
Enables you to define the "IDN String" and the "OPT String" for the
instrument selected with the parameter Language.
Remote command:
:SYSTem:IDENtification on page 443
Set to default
Enables you to reset the *IDN and *OPT strings in user defined mode, see "Mode"
on page 111 .
The default strings vary depending on the selected emulation mode (Language)
IDN String
Indicates the identification string of the instrument when queried with the common
command *IDN?.
To assign a user defined identification string, select Mode "User defined".
Remote command:
*IDN? on page 281
OPT String
Indicates the option string of the instrument as queried with the common command
*OPT?.
General Instrument Settings