[:SOURce<hw>]:FM:EXTernal:DEViation <Deviation>
Sets the deviation of the external frequency modulation signal in Hz. The maximum
deviation depends on the set RF frequency and the selected modulation mode (see
data sheet).
The sum of the deviations of all active frequency modulation signals may not exceed
the total value set with command [:SOURce<hw>]:FM[:DEViation].
Parameters:
<Deviation> float
Range: see data sheet
Increment: 0.01
*RST: 1000
Example:
FM:EXT:DEV 3kHz
sets 3 kHz deviation to the frequency modulation signal.
Manual operation: See "FM Deviation" on page 204
[:SOURce<hw>]:FM:INTernal:DEViation
<Deviation>
Sets the deviation of the internal frequency modulation signal in Hz.
The sum of the deviations of all active frequency modulation signals may not exceed
the total value set with command [:SOURce<hw>]:FM[:DEViation].
Parameters:
<Deviation> float
Range: 0 to dynamic
Increment: 0.01
*RST: 1E3
Example:
FM:INT1:DEV 2E3
sets 2 kHz deviation for the freqeuncy modulation signal.
Manual operation: See "FM Deviation" on page 204
[:SOURce<hw>]:FM:MODE <Mode>
Selects the mode for the frequency modulation.
SOURce Subsystem

Table of Contents

Other manuals for R&S SMB100A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S SMB100A and is the answer not in the manual?

R&S SMB100A Specifications

General IconGeneral
BrandR&S
ModelSMB100A
CategoryInverter
LanguageEnglish