Remote Control Commands
R&S
®
SMB100A
381Operating Manual 1407.0806.32 ─ 21
Parameters:
<Amplitude> float
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model.
The values are given in the data sheet.
Increment: 0.01
*RST: 30
Example:
SOURce:POWer:LIMit:AMPLitude 10
or
:POW:LIM 10
limits the RF level to maximum +10 dBm.
Manual operation: See "Limit - RF Level" on page 147
[:SOURce]:POWer:WIGNore
<State>
Ignores level range warnings.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
POW:WIGN ON
suppresses the level range warnings.
Manual operation: See "Ignore Level Range Warnings" on page 148
[:SOURce<hw>]:POWer:LMODe <LevMode>
Sets the RF level mode.
Parameters:
<LevMode> NORMal | LOWNoise | LOWDistortion
NORMal
The RF signal is output in the standard values of the instrument.
LOWNoise
A very low noise sinewave signal is output.
LOWDistortion
A very pure sinewave signal is output.
*RST: NORMal
Example:
POW:LMODE LOWD
sets the LOWDistortion mode. The instrument reduces distor-
tions of the RF signal to a minimum.
Manual operation: See "RF Mode" on page 147
SOURce Subsystem