Objet30 User Guide 6 Operating and Maintaining the Printer
DOC-34025 Rev. I 6-110
Printer Software
o Run a sanity check of the printer parameters
See "Built-in Tests" on page 6-101.
o On the printer computer desktop, click BackupData.
This saves a backup copy of the Modes and Configs files.
o Restart the printer and server computers.
UV Sensor Calibration
o The UV sensor should be calibrated once every two years. If two years
have passed since the last service visit, contact Stratasys to request
UV sensor calibration.
UPS (Uninterrupted Power Supply) Unit
o Check that the UPS unit is in operating condition.
Refer to the manufacturer's documentation for recommended annual
checks and tests.