Chapter 10
Version 4.0 (1.99) Synthetic Fiber Testing - Page 10-6
fingers or another narrow object such as another fibrocomb until it
can be pulled from the right side and removed.
If the metal length standard will not slide in/out from the track easily,
refer to the following instructions to raise the cabinet top and
insert/remove the comb. Swing the monitor to the side so that the
cabinet top can be lifted. Make sure no loose objects are lying on the
cabinet top. Lift the cabinet top, making sure that the brace on the rod
spring lock is engaged when the cabinet top is raised. The blower will
automatically shut off when the top is lifted.
Slide the length standard along the track into the comb holder, and
position it in front of the clamping mechanism.
Flip the brace from the locked position, and lower the cabinet top to
its original position. The blower will automatically turn on after the
top is closed.
Press ENTER to begin the test. The instrument will perform the test
using the length standard and then display the results indicating
whether the value measured
Passed
or
Failed
.
After the test is complete, raise the cabinet top, and slide the metal
length standard to the right along the track to remove the length
standard. Lower the cabinet top to its original position.