Chapter 10
Version 4.0 (1.99) Synthetic Fiber Testing - Page 10-9
3.5.4 Number of Calibration Tests
Before performing a calibration, the number of repetitions that should be
performed per test should be established. This is done using the
Number of Calibration Tests
item on the SYNTHETIC FIBER TESTING MENU.
Select the
Number of Tests
item using the arrow keys, then press the
ENTER key.
Each calibration will consist of 4, 8 or 12 tests. To change the
number of tests, enter 4, 8 or 12 at the prompt, then press ENTER.
Although performing four (4) tests is an option, it is advised that at least
eight be performed, particularly if there is a great deal of variation in
the fiber or the sampling of the fiber.