Chapter 10
Version 4.0 (1.99) Synthetic Fiber Testing - Page 10-11
3.5.5.1 Synthetic Fiber Standard Values
The values entered on Synthetic Fiber Standard Values screen will be
displayed and used as the Standard Values for Calibration. If no value
is entered for Elongation, Uniformity, or Strength, that property will not
be calibrated.
The Fiber Length value entered here will be displayed on the Calibration
Test Screen as the Len 2 standard value. Len 1 is calculated by
multiplying Len 2 times Uniformity divided by 100.
Change the values by using the arrow keys to highlight the item,
then press the ENTER key. Enter the value at the prompt, and then
press ESCAPE to save the data.