Chapter 10
Version 4.0 (1.99) Synthetic Fiber Testing - Page 10-15
As long as the difference between the Average and the Standard Values
are within Tolerance, the calibration will pass.
• To continue, press ENTER. A printout containing individual test
data, low, high, range, average, and standard values will be
generated.
Note: If you did not enter a calibration value for uniformity or elongation, the
calibration procedure will not calibrate that property.
If one or more of the properties fail calibration, the following message is
displayed.
You have failed Calibration Check.
Press C to continue calibration or E to exit.
• If you press C to Continue, the Calibration screen will re-appear and
the tests must be run again. Only the properties that failed
calibration will be affected by this re-calibration.
• If you press E to Exit, the CALIBRATION MENU will appear.
The calibration process will continue indefinitely as long as calibration
continues to fail. However, no changes are made to the calibration file
until all properties pass.