Chapter 5 95
Setting Measurement Conditions
Using Time as Sweep Parameter (Zero Span Sweep)
5. Setting Measurement
Conditions
[Utility] - MKR X-AXIS [ ] - TIME keys.
Sweep parameters other than frequency can be selected in Step 3. In such a case, the fixed
value for the signal source is reflected in the selected parameter in Steps 5 and 6.
Settings for signal sources other than the fixed value in Steps 5 and 6 can be made in the
signal source menu, which can be accessed by pressing the
[Source] key.
The zero span sweep can also be achieved with the
[Start] and [Stop] keys to specify the
sweep start value and sweep stop value as the same value.
Figure 5-10 Sweep span value = 20 kHz