Appendix B 383
Key Definitions
Softkeys displayed by pressing the [Sweep] key
B. Key Definitions
Softkeys displayed by pressing the [Sweep] key
Table B-43 Sweep (Sweep) menu
Key operation: [Sweep]
Softkey Function GPIB command
PARAMETER [ ] ä Displays Table B-44, “Sweep parameter (Sweep Param) menu,” on page 383. In
[ ] of the softkey label, the currently selected sweep parameter is displayed
(frequency sweep [FREQ], signal level sweep [OLEV], or dc bias sweep
[DCB]).
(N/A)
NUMBER OF POINTS Sets the number of measurement points per sweep (number of points, NOP)
when the sweep type is not LIST (LINEAR or LOG). Enter an integer between 2
and 801. You set the number of points for list sweep in the list sweep table.
POIN <value>
DIRECTION [ ] Switches the sweep direction between upward [UP] and downward [DOWN].
The upward direction is the direction in which the sweep parameter value
increases; the downward direction is the direction in which the sweep parameter
value decreases.
SWED {UP|DOWN}
MANUAL SWP on OFF Toggles on and off the manual sweep function. Immediately after pressing the
MANUAL SWP on OFF key, you can freely move to one of the specified sweep
points (1–number of points) with the rotary knob, step keys ([], [¯]), or numeric
entry keys and perform measurements at the point. If the manual sweep function
is set to on, you can also enter sweep points (1–number of points) with the
[sweep] key.
MANS <value>
TIME ä Displays Table B-45, “Sweep time (Sweep Time) menu,” on page 384. (N/A)
TYPE [ ] ä Displays Table B-46, “Sweep type (Sweep Type) menu,” on page 384. In [ ] of
the softkey label, the currently selected sweep type is displayed (linear sweep
[LINEAR], log sweep [LOG], or list sweep [LIST]).
(N/A)
EDIT LIST ä Displays Table B-47, “List sweep (List Sweep) menu,” on page 384. EDITLIST
LIST SPAN [ ] Switches the list span (configuration of the horizontal axis of the graph in list
sweep) between single span [SINGLE] or segment span [SEGMENT]. Single
span is a method to set the linear frequency axis as a horizontal axis whose left
edge is the minimum frequency and whose right edge is the maximum frequency
in the entire list sweep table. On the other hand, segment span is a method to set,
as the horizontal axis of the graph, an axis in which the measurement points of
each segment are equally positioned in the measurement order in the entire table
from the left edge to right edge of the graph (measurement number order axis). If
COMPLEX Z-Y has been selected as measurement parameter, there is no sweep
parameter axis on the graph and therefore this configuration is invalid.
LISPAN
{SINGLE|SEGMENT}
Table B-44 Sweep parameter (Sweep Param) menu
Key operation: [Sweep] - PARAMETER [ ]
Softkey Function GPIB command
FREQ
|
Selects frequency as the sweep parameter. SWPP FREQ
OSC LEVEL
|
Selects oscillator level as the sweep parameter. SWPP OLEV
DC BIAS Selects dc bias as the sweep parameter. SWPP DCB

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals