Chapter 4 67
Preparation of Measurement Accessories
Adapter Setting
4. Preparation of
Measurement Accessories
Adapter Setting
You must ensure that the adapter setting is properly set after you finish connecting
accessories to the four-terminal pair port of the Agilent 4294A and before you start
measurement. If required, change it to the correct adapter setting. The adapter setting
includes adapter selection and the data acquisition process, which is called adapter setup.
Table 4-2 lists the device connections and corresponding adapter settings.
Table 4-2 Device Connections and Adapter Settings
Device Connection Accessory
Adapter
Selection
Adapter Setup (data measurement)
(Ö: required; blank: not required)
Phase
1
OPEN SHORT LOAD
A. Connect a direct
connection type test fixture
to the four-terminal pair port
of the Agilent
4294A and
connect the device to the
fixture contacts.
16034E
16034G
16034H
16044A
16047A
16047C
16047D
16047E
16065A
2
16089A
2
16089B
2
16089C
2
16089D
2
16089E
2
16314A
No Adapter (
NONE)
B. Extend the four-terminal
pair port of the Agilent
4294A with a 1-meter-long
four-terminal pair
configuration and connect
the device.
16048G and test fixture
16048G and 16452A
Four-terminal pair,
1 meter (
4TP 1M)
ÖÖ
16334A Four-terminal pair,
1 meter (
4TP 1M)
ÖÖ
16451B Four-terminal pair,
1 meter (
4TP 1M)
Ö
C. Extend the four-terminal
pair port of the Agilent
4294A with a 2-meter-long
four-terminal pair
configuration and connect
the device.
16048H and test fixture
16048H and 16452A
Four-terminal pair,
2 meters (
4TP 2M)
ÖÖ
D. Connect the 42942A
directly to the four-terminal
pair port of the Agilent
4294A, connect a dedicated
test fixture to the 7-mm port,
and then connect the device
to the fixture contacts.
42942A and 16092A
42942A and 16093A
42942A and 16093B
42942A and 16191A
42942A and 16192A
42942A and 16193A
42942A and 16454A
42942A Terminal
Adapter (
7mm
42942A
)
ÖÖÖÖ

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals