EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>4294A

Agilent Technologies 4294A User Manual

Agilent Technologies 4294A
469 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #321 background imageLoading...
Page #321 background image
Chapter 9 313
Setting/Using Control and Management Functions
Setting/Checking the GPIB
9. Setting/Using Control and
Management Functions
Setting/Checking the GPIB
This section describes how to set or check the interface necessary to use the Agilent
4294A's GPIB (General Purpose Interface Bus). For more information on automatic
measurements using the GPIB, refer to Programming Manual.
Switching between the System Controller Mode and Addressable-only
Mode
When you use the Agilent 4294A in an GPIB system, you have to specify whether to use it
as the system controller (system controller mode) or in the state where addressing is always
made from other devices (addressable-only mode). Follow these steps to switch between
the system controller mode and the addressable-only mode.
Step 1. Press the
[Local] key to display the Local menu.
Step 2. Select the system controller mode or addressable-only mode with one of the following
keys.
NOTE There should only be one system controller on a GPIB system. If there is another system
controller, set this instrument to the addressable-only mode. If you try to set the Agilent
4294A as the system controller using the above key while there is another system
controller, the “CAUTION: CAN'T CHANGE - ANOTHOR CONROLLER ON BUS
error message is displayed.
To pass the GPIB control from the controller to this instrument, set this instrument to the
addressable-only mode in advance.
Setting/Checking the GPIB address
Follow these steps to set or check the GPIB address of this instrument.
Step 1. Press the
[Local] key to display the Local menu.
Step 2. Press the
ADDRESS: HP-IB key. In the set parameter value field in the upper left area of
the screen, the currently set GPIB address is displayed.
Step 3. Use the keys or rotary knob of the ENTRY block in one of the following ways to specify
the GPIB address.
Enter the desired value with the numeric keys (
[0] to [9]) and then press the [´1] unit
key.
Turn the rotary knob (
m
mm
m) until the desired value is set.
Press the step keys ([][¯]) to set the desired value.
The GPIB address can be set to an integer ranging from 0 through 30.
GPIB mode Key operation
system controller mode
SYSTEM CONTROLLER
addressable-only mode ADDRESSABLE ONLY MODE

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals