EasyManuals Logo

Agilent Technologies 4294A User Manual

Agilent Technologies 4294A
469 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #141 background imageLoading...
Page #141 background image
Chapter 6 133
Calibration
A. Calibration When Using Direct Connection Type Test Fixture
6. Calibration
A. Calibration When Using Direct Connection Type Test
Fixture
This section describes the calibration procedures to follow when a direct connection type
test fixture is connected to the four-terminal pair port of the Agilent 4294A and a device is
connected to the test fixture contacts.
Figure 6-1 Calibration When Direct Connection Type Test Fixture is Used
Direct connection type test fixtures include:
16034E Chip component test fixture
16034G Chip component test fixture
16034H Chip component test fixture
16044A Chip component test fixture
16047A Lead component test fixture
16047C Lead component test fixture
16047D Lead component test fixture
16047E Lead component test fixture
16314A Balanced/Unbalanced 4-Terminal Converter
Although the following test fixtures have cables, their calibration must be performed in the
same way as a direct connection type test fixture:
16065A External DC Bias Fixture
16089A Kelvin Clip Lead

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals