Chapter 8 275
Analysis and Processing of Result
Save and Recall the Agilent 4294A Internal Data
8. Analysis and Processing
of Result
NOTE There is no need to attach an extension to the file name.
Step 11. After entering the file name into the character string input line, press the
done key or [´1]
key to save the file. To suspend saving, press the
cancel key. In both cases, you will return
to the Data Save menu.
NOTE Steps 10 and 11 can be executed from the external keyboard connected to the rear panel.
After entering the file name, press the
[Enter] key of the keyboard to save the file.
The extension [.DAT] is automatically attached to the name of the file to be saved when
saved in binary format, and the extension [.TXT] is automatically attached to it when saved
in text (ASCII) format.
It is not possible to later recall the calibration data and trace data saved in text (ASCII)
format for reuse by the Agilent 4294A.
The calibration data and trace data saved in binary format does not contain setting
information such as the sweep range or number of sweep points. Accordingly, to recall the
data and use it correctly on the Agilent 4294A, save the setting state using the STATE
saving function.
Using files saved in the text (ASCII) format with the data save function
You can read data in files saved in the text (ASCII) format as described in “Save the
calibration data and trace data (Data Save)” on page 273 using software such as a text
editor on your PC (personal computer). You can also import it into the spreadsheet
software for further applications. Table 8-5 shows an example of data saved in the text
(ASCII) format.
BACK SPACE or [Back Space] Returns the cursor (_)of the character string
input line by one character to the left and
erases the character in that position.
CLEAR NAME Erases the entire the character string entered in
the character string input line.
Key Function

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals