Chapter 6 131
Calibration
Selecting Appropriate Calibration Method
6. Calibration
Refer to the following sections for the calibration procedure of each device connection:
“A. Calibration When Using Direct Connection Type Test Fixture” on page 133
“B. Calibration for Four-Terminal Pair, 1-m Extension” on page 135
“C. Calibration for Four-Terminal Pair, 2-m Extension” on page 137
“D. Calibration When an Exclusive Fixture is Connected to the 42942A” on page 139
“E. Calibration When the 7-mm Port of the 42942A is Extended” on page 141
“F. Calibration When a Probe Adapter is Connected to the 42941A” on page 143
“G. Calibration When the 3.5-mm Port of the 42941A is Extended” on page 145
For the three calibration methods, refer to the following sections:
“User Calibration” on page 147
“Port Extension Compensation” on page 150
“Fixture Compensation” on page 151
D. Connect the 42942A
directly to the four-terminal
pair port of the 4294A, connect
a dedicated test fixture to the
7-mm port, and then connect
the device to the fixture
contacts.
42942A
Terminal
Adapter (
7mm
42942A
)
42942A and 16092A
42942A and 16093A
42942A and 16093B
42942A and 16191A
42942A and 16192A
42942A and 16193A
Ö
(perform when
user-defined
calibration kit is
used)
Open/Short
42942A and 16454A Short
E. Connect the 42942A directly
to the four-terminal pair port of
the 4294A, connect a coaxial
cable to the 7-mm port, connect
a test fixture, and then connect
the device to the fixture
contacts.
42942A
Terminal
Adapter (
7mm
42942A
)
42942A, coaxial cable,
and test fixture
Ö
(use Agilent's
calibration kit or
user-defined
calibration kit)
Ö Same as the fixtures
are directly
connected to the
terminal adapter.
F. Connect the 42941A directly
to the four-terminal pair port of
the 4294A, connect a dedicated
probe adapter to the 3.5-mm
port, and then connect the
device to the adapter contacts.
42941A
Impedance
Probe (
PROBE
42941A
)
42941A and dedicated
probe adapter
Ö
(perform when
user-defined
calibration kit is
used)
Open/Short
G. Connect the 42941A directly
to the four-terminal pair port of
the 4294A, connect a coaxial
cable to the 3.5-mm port,
connect a test fixture, and then
connect the device to the
fixture contacts
42941A
Impedance
Probe (
PROBE
42941A
)
42941A and coaxial
cable and test fixture
Ö
(use Agilent's
calibration kit or
user-defined
calibration kit)
Ö Required
compensations
differ from one test
fixture to anather.
1. Although it has a cable, it falls into the category of direct connection type test fixture for adapter setup.
Table 6-1 Device Connection and Corresponding Adapter Selection and Calibration
Device Connection
Adapter
Selection
Accessory
Calibration (Ö: required, blank: not required)
User
Calibration
Port Extension
Compensation
Fixture
Compensation

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals