Chapter 4 77
Preparation of Measurement Accessories
Adapter Setting
4. Preparation of
Measurement Accessories
Figure 4-6 Connecting SHORT (0 W) to the 42942A
Step 9. Press the
SHORT [-] key to start the short data measurement.
NOTE When the short data measurement is completed, the softkey label changes to
SHORT [DONE].
Step 10. Remove the SHORT (0 W) standard from the 7-mm port of the 42942A. Then connect the
LOAD (50 W) standard to the 7-mm port (the LOAD standard is furnished with the
42942A Terminal Adapter).