5.2 PPG/ED Panel
5-23
5
How to Operate BERT
Measurement method
The ED Result display can be refreshed either in real time (approx. 0.1 s
intervals) or when the progress bar reaches 100%. The display method is
set at Current of Gating.
On: Updates measurement results in real time.
Off: Updates measurement results when progress reaches 100% for either
Single or Repeat Gating Cycle.
Updates measurement results when measurement stops for Untimed
Gating Cycle.
The setting procedure of the measurement conditions is as follows:
1. Click the Gating Cycle button and select the measurement method as
follows:
Repeat
Single
Untimed
2. When Gating Cycle is Repeat or Single, click the text box under the
Gating Cycle and enter the numeric value.
The measurement cycle can be set from 1 second to 9 days 23 hours
59 minutes 59 seconds.
3. Click Current to set the timing for the measurement result.
On: Updates measurement results per 100 ms.
Off: Displays measurement results per measurement results
or when measurement stops
4. Click Start / Stop.
The right indicator of the button is changed to green.
Measure is displayed at the status display.
The measurement processing rate is displayed.
When pattern synchronization is obtained, the SYNC Loss display is
erased.
When Gating Cycle is set to Untimed, the measurement cycle progress is
displayed every 5 seconds.