22.)If the slider is too high, turn the Volume Dial on the receiver slightly CW
23.)If the slider is too low, turn the Volume Dial on the receiver slightly CCW
24.)Repeat the Microphone Setup and turn the Volume Dial until the slider is in a good spot.
25.)The receiver is now set up correctly…reassemble the covers.
Figure 8-205 Volume Slider
Bottom of slider should be in this range
Bottom of slider can not be touching
the bottom tick mark
Top of slider can not be touching the
top tick mark
Troubleshooting: If slider
moves to the top and stays
there, and you have to keep
reading the paragraph, no
audio is entering the USB
Adapter from the Receiver.