REMOTE CONTROL
259
VAL1?
Returns the 1
st
display value.
Example: SAMP:COUN 5
VAL1?
>+0.33452387E-4
>+0.32897125E -4
> etc, for 5 counts.
Queries 5 counts of stored samples from the 1
st
display.
VAL2?
Returns the 2
nd
display value.
Example: SAMP:COUN 5
VAL2?
>+0.38954687E -4
>+0.32764551E -4
> etc, for 5 counts.
Queries 5 counts of stored samples from the 2
nd
display.
ROUTe:TERMinate?
Indicates which input terminals are selected on the GDM-9061 front panel
Front/Rear switch. This
switch is not programmable; this query reports the position of the switch,
but cannot change it.
Return parameter: FRON | REAR
On the GDM-9060, this query always returns FRON.
TIME:SYNC:SERVer
Sets the server source for time sync of the 2nd group.
Parameter: ”<server>, max length = 22 characters.
Example: TIME:SYNC:SERV time-nv.nist.gov
TIME:SYNC:SERVer?
Returns the server source for time sync of the 2nd group.
Return parameter: <server>, Ex: time-nv.nist.gov
CALCulate Commands
CALCulate:CLEar[:IMMediate]
Clears all of the compare results, statistic calculation value, histogram
calculation value, and measurement value.
Parameter: <None>
Example: CALC:CLE:IMM
CALCulate:DATA?
Returns uncalculated original measurement.