REMOTE CONTROL
263
CALCulate:DB:REFerence:METHod?
Returns the unit of reference value from the dB function.
Return parameter: Voltage | dBm
CALCulate:DBM:REFerence
Sets the resistance value for the dBm function.
Parameter: <NR1> (2, 4, 8, 16, 50, 75, 93, 110, 124, 125, 135, 150, 250, 300,
500, 600, 800, 900, 1000, 1200, 8000) | MIN | MAX | DEF
Example: CALC:DBM:REF MAX
Sets the resistance value for dBm measurements to the maximum allowed.
CALCulate:DBM:REFerence?
Returns the resistance value from the dBm function.
Return parameter: <NRf>
CALCulate:SCALe:REFerence:AUTO
Sets the first measurement as the reference value.
Parameter: 0 | 1 | ON | OFF
Example: CALC:SCAL:REF:AUTO ON
Sets the reference value auto setting on for dB measurement.
CALCulate:SCALe:REFerence:AUTO?
Returns the auto setting status of the dB function.
Return parameter: 0 | 1, 1=ON, 0=OFF
CALCulate:MATH:MMFactor
Sets the scale factor M for math measurement.
Parameter: <NRf> | MIN | MAX | DEF
Example: CALC:MATH:MMF MIN
Sets the scale factor M to the minimum allowed value.
CALCulate:MATH:MMFactor?
Returns the scale factor M used in the math measurement.
Return parameter: <NRf>
CALCulate:MATH:MBFactor
Sets the offset factor B for math measurement.
Parameter: <NRf> | MIN | MAX | DEF
Example: CALC:MATH:MBF MIN
Sets the offset factor B to the minimum allowed value.
CALCulate:MATH:MBFactor?
Returns the offset factor B used in the math measurement.
Return parameter: <NRf>