REMOTE CONTROL
279
[SENSe:]DIGital:SHIFt?
Returns the digital shift function status.
Return parameter: 0 | 1 ,1=AUTO, 0=User selected
[SENSe:]UNIT
Sets the temperature unit.
Parameter: C | F | K
Example: SENS:UNIT C
Sets the temperature unit to ˚C.
[SENSe:]UNIT?
Returns the temperature unit.
Return parameter: C | F | K
SENSe AVERage Commands
[SENSe:]AVERage:COUNt[X]
Sets the digital filter count, which X = 1 indicate 1
st
display, X = 2 indicate
2
nd
display.
Parameter: <NR1> (2 ~ 100) | MIN | MAX | DEF
Example: SENS:AVER:COUN2 100
Sets 2
nd
display digital filter count number to 100.
[SENSe:]AVERage:COUNt[X]?
Returns the digital filter count.
Return parameter: <NR1>, Ex: +002
[SENSe:]AVERage:STATe[X]
Turns the digital filter function On/Off, which X = 1 indicate 1
st
display, X =
2 indicate 2
nd
display.
Parameter: 0 | 1 | ON | OFF
Example: SENS:AVER:STAT ON
Turns 1
st
display digital filter function on.
●If NPLC >= 7.2k/s, the filter function will be disabled.
[SENSe:]AVERage:STATe[X]?
Returns the state of the digital filter function (on or off).
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]AVERage:TCONtrol[X]
Selects the digital filter type, which X = 1 indicate 1
st
display, X = 2 indicate
2
nd
display.
Parameter: MOV | REP
Example: SENS:AVER:TCON MOV
Sets 1
st
display digital filter type to the moving filter.