REMOTE CONTROL
291
[SENSe:]CURRent:AC:TRIGger:DELay?
Returns the trigger delay time in seconds of AC Current measurement.
Return parameter: <NRf>
SENSe RESistance Commands
[SENSe:]RESistance:NPLCycles
Sets the integration time for 2-wire resistance measurement in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.006 | 0.0083 | 0.0125 | 0.025 | 0.05 | 0.15 | 0.6 | 1 | 3
| 12).
Parameter: NPLCycles(<NRf> | MIN | MAX | DEF)
Example: SENS:RES:NPLC MIN
Sets the integration time to 0.006 PLCs for 2-wire resistance measurement.
[SENSe:]RESistance:NPLCycles?
Returns the integration time for 2-wire resistance measurement in PLCs
(power line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.006 | 0.0083 | 0.0125 | 0.025 | 0.05 | 0.15 | 0.6 | 1 | 3 | 12
[SENSe:]RESistance:NULL[:STATe]
Sets the relative function on/off for 2-wire resistance measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:RES:NULL:STAT ON
Turns the relative function on for 2-wire resistance measurement.
[SENSe:]RESistance:NULL[:STATe]?
Returns the relative function state of 2-wire resistance measurement.
Return parameter: 0|1, 0=OFF, 1=ON
[SENSe:]RESistance:NULL:VALue
Sets the relative value for 2-wire resistance measurement.
Parameter: <NRf> (-120.0~120.0 MΩ) | MIN | MAX | DEF
Example: SENS:RES:NULL:VAL 2
Sets the relative value to 2Ω for 2-wire resistance measurements.
[SENSe:]RESistance:NULL:VALue?
Returns the current relative value of 2-wire resistance measurement.