GDM-906X Series User Manual
292
[SENSe:]RESistance:NULL:VALue:AUTO
Sets the relative value auto on/off for 2-wire resistance measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:RES:NULL:STAT ON
SENS:RES:NULL:VAL:AUTO OFF
The unit automatically sets the 1st count of measurement as null value.
[SENSe:]RESistance:NULL:VALue:AUTO?
Returns the null value auto state of 2-wire resistance measurement.
[SENSe:]RESistance:RANGe
Sets the 2-wire resistance measurement range.
Parameter: Range(<NRf> | MIN | MAX | DEF)
Example: SENS:RES:RANG 1000
Sets the 2-wire resistance range to 1kΩ.
[SENSe:]RESistance:RANGe?
Returns the 2-wire resistance measurement range.
[SENSe:]RESistance:RANGe:AUTO
Sets the 2-wire resistance Auto-range settings on, off or once.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:RES:RANG:AUTO ON
Turns Auto-range on for 2-wire resistance measurement.
[SENSe:]RESistance:RANGe:AUTO?
Returns the 2-wire resistance Auto-range setting.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]RESistance:RESolution
Sets the 2-wire resistance measurement resolution. The resolution depends
on the rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX | DEF)
Example: SENS:RES:RES 0.01
Sets the 2-wire resistance resolution to 0.01
[SENSe:]RESistance:RESolution?
Returns the 2-wire resistance resolution.
[SENSe:]RESistance:TRIGger:DELay
Sets the trigger delay time that minimum step is microseconds of 2-wire
resistance measurement.
Parameter: <NRf>(0 ~ 3600 s) | MIN | MAX | DEF
Example: SENS:FRES:TRIG:DEL DEF
Sets the trigger delay time to 1s of 2-wire resistance measurement.