105
Chapter 5, AF Analyzer Screen
Block Diagram
Detector
This setting selects the type of detector used when measuring and displaying AF
signal levels.
Detector Types
RMS displays the RMS value of signals.
displays the RMS value of a signal multiplied by .
Pk+ displays the positive peak value.
Pk displays the negative peak value.
Pk±/2 adds the positive and negative peak values, and divides the sum by 2.
Pk±Max compares the positive and negative peaks and displays the greater value
(polarity is not indicated).
Pk+ Hold displays and holds the positive peak value until the measurement is reset.
To reset, press the MEAS RESET key, select a different detector, or re-select the same
detector.
Pk Hold displays and holds the negative peak value until the measurement is reset.
To reset, press the MEAS RESET key, select a different detector, or re-select the same
detector.
Pk±/2 Hold divides the sum of the positive and negative peak values by 2, and
displays the value until the measurement is reset. To reset, press the MEAS RESET
key, select a different detector, or re-select the same detector.
Pk±Mx Hold compares the positive and negative peaks and displays the greater value
until the measurement is reset. To reset, press the MEAS RESET key, select a different
detector, or re-select the same detector.
See Also
"Interaction Between Screens" on page 54