Chapter 24, Connector, Key, and Knob Descriptions
Connector Descriptions
AUDIO MONITOR OUTPUT
This connector provides an external output from various tap points in the AF
Analyzer.
Operating Considerations
The Scope To field in the AF ANALYZER screen determines the source of this
signal.
The level is not affected by the front-panel VOLUME knob.
Output impedance <1 k
Ω
See Also
Chapter 5, "AF Analyzer Screen"
AUDIO OUT
This port is used to output signals from AF Generators 1 and 2, including the
encoder functions.
Operating Considerations
The output level is set by the AF Generators and is not affected by the front-panel
VOLUME control.
Output impedance <1
Ω
Maximum output current = 20 mA peak
Maximum reverse voltage = 12 V peak
AC/DC coupling is selected using the
Audio Out field. This field is available in
the DUPLEX TEST screen, RF GENERATOR screen, and various encoder
modes.