536
S:\HP892XX\USRGUIDE\manual\testscrn.fb
Chapter 22, Tests Screen
TESTS (Execution Conditions)
Autostart Test Procedure on Power-Up
When set to On, this field automatically loads and runs the specified procedure
whenever the Test Set is turned on. If this field is set to
Off or the specified
procedure cannot be loaded for any reason (such as, memory card not inserted,
disk drive not connected), the Test Set will default to its normal power-on state.
Continue
This field is used to restart a paused test.
If Unit-Under-Test Fails
This field selects what happens if the measured results do not meet criteria for
passing the current test. The underlined entry is the active choice.
• Continue continues with testing even though a test in the test sequence fails to meet
its test specification limits. When this occurs, an error is listed on the test-results print-
out, is displayed on the CRT, or both.
• Stop stops the test when a failure occurs and requires operator intervention before test-
ing proceeds.
Main Menu
Selecting this field returns you to the TESTS (Main Menu) screen.
Output Heading
This field allow you to enter a heading for the test results printout (or CRT
display).
Output Results To
This field selects where test results are output. The underlined entry is the active
choice.
• CRT displays test results on the Test Set’s CRT.
• Printer displays test results on the Test Set’s CRT and outputs them to a printer.
(The printer information must first be entered on the TESTS (Printer Setup) screen.)