537
Chapter 22, Tests Screen
TESTS (Execution Conditions)
Output Results For
This field selects which test results are output. The underlined entry is the active
choice.
• All outputs all test results on to the location selected in the Output Results To:
field.
• Failures outputs only the results of tests which fail.
Run Test
Selecting this field loads and runs the test chosen in the Select Procedure
Filename:
field on the TESTS (Main Menu) screen.
Test Procedure Run Mode
This field selects how the test will be run. The underlined entry is the active
choice.
• Continuous runs the tests in the test sequence one after another without stopping.
Testing pauses only if the operator is required to interact with the UUT or Test Set. In-
teractions such as changing UUT channels and setting squelch and audio levels cause
testing to pause. Selecting Stop in the If Unit-Under-Test Fails: field may
also stop a test.
• Single Step runs the tests in the test sequence one at a time. The operator is prompt-
ed to press Continue to proceed with testing.