EasyManua.ls Logo

HP 8920A - Sinad; Snr; SNR Operation

HP 8920A
658 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
110
S:\HP892XX\USRGUIDE\manual\afascrn.fb
Chapter 5, AF Analyzer Screen
Block Diagram
SINAD
This measurement field is used to select and display any one of the following
measurements:
SINAD
Distortion
SNR (Signal to Noise Ratio)
AF Frequency
DC Level
Current - DC only
1
Operating Considerations
This is a priority control field. Accessing the RX TEST or TX TEST screen
overrides any changes made to this field in other screens.
SNR Operation
Selecting SNR turns off the other audio measurement.
The RF Generator and AFGen1 must be set up to provide the radio’s carrier. (AFGen1
is automatically turned on and off repeatedly during this measurement.)
AFGen2 must be turned off.
The radio’s receiver’s audio output must be connected to the AUDIO IN port (set the
AF Anl In field to Audio In).
See Also
"Interaction Between Screens" on page 54
1. HP 8920A requires Current Measurement option
SNR
The SNR field is found by selecting the SINAD field on the Test Set display of the
AF ANALYZER screen.This field displays the signal to noise ratio of the input
audio signal.