Chapter 3, Operating Overview
Saving and Recalling Instrument Setups
To Clear All Save Registers
1. Press RECALL.
2. Use the knob to position the cursor in front of the *Clr All* entry in the Recall
menu at the bottom-right of the screen.
3. Press the knob or press ENTER. A prompt appears at the top of the screen to verify that
you want to clear all registers.
4. Press YES.
Register Names
You can use any number, letter, or combination of numbers and letters as a name
for storing instrument settings. For instance; if you want to save a setup for testing
a “Vulcan7” radio, you can save the setting as “VULCAN7”.
Two register names are reserved for special purposes: POWERON and BASE.
POWERON Settings
When the Test Set is turned on, it uses a set of instrument setup parameters
specified at the time of manufacture. You can have the instrument power up in a
different state by making the desired changes to the original settings, and then
saving them using the name POWERON.
The next time the instrument is turned on, the instrument returns to the state
present when you saved the POWERON setting. For instance; if the
OSCILLOSCOPE screen was displayed when POWERON was saved, it is the
screen that is displayed when you turn the instrument on.