S:\HP892XX\USRGUIDE\manual\dxtscrn.fb
Chapter 9, Duplex Test Screen
Field Descriptions
AFGen1 Freq
This field sets the frequency for the first audio frequency sinewave generator.
AFGen1 To
This field is used to set two values for the first audio frequency generator:
• The upper field determines whether the AF signal modulates the RF Generator, or is
output through the AUDIO OUT connector.
• The lower field sets the amplitude (including
Off
). The AUDIO OUT level is always
in volts RMS.
Operating Considerations
This is a priority control field. Accessing the RX TEST or TX TEST screen
overrides any changes made to this field in other screens.
See Also
"Interaction Between Screens" on page 54