478
S:\HP892XX\USRGUIDE\manual\rfanscrn.fb
Chapter 17, RF Analyzer Screen
Field Descriptions
Trying to read a squelched measurement using HP-IB will cause your program to
halt until the squelch is either turned down, a measurement is made, or until a
program time-out aborts the measurement process.
SINAD
This is the default measurement for this field. Selecting this field displays the
following measurement choices:
SINAD
Distortion
SNR (Signal to Noise Ratio)
AF Frequency
DC Level
Current - DC only
1
Operating Considerations
This is a priority control field. Accessing the RX TEST or TX TEST screen
overrides any changes made to this field in other screens.
SNR Operation
• Selecting SNR turns off the other audio measurements.
• The RF Generator and AFGen1 must be set up to provide the radio’s carrier. (AFGen1
is automatically turned on and off repeatedly during this measurement.)
• AFGen2 must be turned off.
• The radio’s receiver’s audio output must be connected to the AUDIO IN port (set the
AF Anl in field to Audio In.
See Also
"Interaction Between Screens" on page 54
"Displaying Measurements" on page 57
1. The Test Set requires Current Measurement option.