510
S:\HP892XX\USRGUIDE\manual\specscrn.fb
Chapter 21, Spectrum Analyzer Screen
Field Descriptions
Output Port
This field selects the output port for the RF Generator.
Operating Considerations
Maximum signal levels at each port are printed on the front panel.
CAUTION: Applying reverse RF power to the DUPLEX OUT connector can damage the instrument. (A
message is displayed when an overpower conditions occurs.) Whenever possible when testing
transceivers, use the RF IN/OUT connector to prevent damage from accidental transmitter
keying.
If a reverse power condition triggers the internal protection circuit, remove the
reverse power signal and press the MEAS RESET key or turn the Test Set off
and on to reset it.
RF Gen Freq
This field sets the RF Generator’s frequency.