3-31
Troubleshooting
ESG Series Signal Generator Instrument-Specific Error Messages
Syntax error;Bad HP compatibility language
token <token>.
Indicates that, in HP 8656/57 compatibility mode, a known
command or termination specifier was received when it was not
expected. For example, a termination specifier was received
with no currently active function.
-213
Init ignored;Unable to sweep due to sweep being
in an error state. The sweep error should be
fixed.
Indicates that the number of list, power, and/or dwell points are
in conflict, or a serious system error has occurred in list/sweep.
A previous error report should have described the error that is
stalling list/sweep.
Init ignored;Cannot initiate sweep in manual
mode.
Indicates that the manual mode is on and therefore the
instrument cannot sweep.
Init ignored;Sweep is already initiated.
Indicates that the list/sweep is currently initiated and sweeping,
therefore the command is not legal according to SCPI.
Init ignored;Sweep is already continuously
initiated.
Indicates that the list/sweep is continuously initiated and
sweeping, therefore the command is not legal according to
SCPI.
-221
Settings conflict;Frequency list and dwell list
are of unequal size. Set one list equal to size
one, or make their sizes equal.
Indicates that the frequency list has more than one element and
the dwell list has more than one element, and they are not of
equal size. If any of the frequency, power, or dwell lists have
more than one element, they must all have the same number of
elements. A list of a single element is the same as a list of equal
size with the single element repeated the necessary number of
times.