3-32
Troubleshooting
ESG Series Signal Generator Instrument-Specific Error Messages
Settings conflict;Frequency list and power list
are of unequal size. Turn one list off, set one
to size one, or make their sizes equal.
Indicates that the frequency list has more than one element and
the power list has more than one element, and they are not of
equal size. If any of the frequency, power, or dwell lists have
more than one element, they must all have the same number of
elements. A list of a single element is the same as a list of equal
size with the single element repeated the necessary number of
times.
Settings conflict;Power list and dwell list are
of unequal size. Set one to size one, or make
their sizes equal.
Indicates that the dwell list has more than one element and the
power list has more than one element, and they are not of equal
size. If any of the frequency, power, or dwell lists have more
than one element, they must all have the same number of
elements. A list of a single element is the same as a list of equal
size with the single element repeated the necessary number of
times.
Settings conflict;The selected external
trigger setting conflicts with the previous
setting.
Indicates that the external trigger has been set to positive edge
for one trigger source and negative edge for another trigger
source.