EasyManua.ls Logo

HP ESG Series - Page 76

HP ESG Series
348 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-33
Troubleshooting
ESG Series Signal Generator Instrument-Specific Error Messages
Settings conflict;FM2/PM2 value set greater
than FM1/PM1 value. FM1/PM1 changed to match
FM2/PM2 value.
The deviation of FM2/PM2 must always be less than or equal
to the deviation settings for FM1/PM1. This error will be
reported to the queue when FM1/PM1 is enabled and
FM2/PM2 is also enabled and an adjustment to either
FM2/PM2 deviation causes the FM2 or PM2 deviation to be
greater than the FM1 or PM1 deviation. It will also be reported
when FM2/PM2 is being turned on, and the last FM1/PM1
deviation setting is less than the current FM2/PM2 deviation
setting. In both cases the FM1/PM1 deviation will be adjusted
to match the FM2/PM2 deviation.
Settings conflict;FM1/PM1 value set less than
FM2/PM2 value. FM2/PM2 changed to match FM1/PM1
value.
The deviation of FM2/PM2 must always be less than or equal
to the deviation settings for FM1/PM1. This error will be
reported to the queue when FM2/PM2 is enabled and
FM1/PM1 is also enabled and an adjustment to either
FM1/PM1 deviation causes the FM1 or PM1 deviation to be
less than the FM2 or PM2 deviation. It will also be reported
when FM1/PM1 is being turned on, and the last FM2/PM2
deviation setting is greater than the current FM1/PM1 deviation
setting. In both cases the FM2/PM2 deviation will be adjusted
to match the FM1/PM1 deviation.
Settings conflict;Enabled mod source conflicts
with previously enabled mod source. Previous
mod disabled.
The signal generator has three sources: INT, EXT1, and EXT2
that are shared by the FM1/PM1, AM1/AM2, FM2/PM2, pulse
(INT and EXT2), and burst envelope (EXT1 only). Each source
can only be used by one of the modulations at a time. If a
source is being used by an active modulation, and a request for
the source is made by another modulation, the first modulation
will be turned off, the second modulation will be turned on.