3-34
Troubleshooting
ESG Series Signal Generator Instrument-Specific Error Messages
Settings conflict;FM & PM not allowed.
Indicates that there is a hardware conflict between FM and PM.
The most recently requested modulation will be turned on, the
previous modulation will be turned off.
Settings conflict;Pattern repeat is changed to
continuous because data source is external.
Indicates that, while in non-bursted data generation, Pattern
Repeat was in Single mode and data source was selected to be
External. For non-bursted data generation using an external
data source, Pattern Repeat must be in Continuous mode. To
continue data transmission, Pattern repeat has been changed to
Continuous mode.
-222
Data out of range;value clipped to lower limit.
Indicates that an input value is below the minimum value
allowed. Examples are: frequency setting, reference, or offset;
output power; power reference and offset; modulation depth,
deviation, or modulation source frequency; number of points
and start/stop values for list mode; sequence or register values
(save/recall); dwell time.
Data out of range;value clipped to upper limit.
Indicates that an input value is above the maximum value
allowed. Examples are: frequency setting, reference, or offset;
output power; power reference and offset; modulation depth,
deviation, or modulation source frequency; number of points
and start/stop values for list mode; sequence or register values
(save/recall); dwell time.
Data out of range;Synthesizer: Frequency out of
bounds.
Indicates that the instrument received an internal request for a
frequency outside of its supported frequency range. Report the
circumstances to the factory.

Table of Contents