3-35
Troubleshooting
ESG Series Signal Generator Instrument-Specific Error Messages
Data out of range;Manual point exceeds list
sizes. Limiting to maximum point.
Indicates that the sweep/list manual point has been reassigned
to a smaller number value due to the longest list decreasing in
size or being turned off. Its new value is the length of the
longest enabled list (frequency or power).
Data out of range;Manual point exceeds
frequency list size. Limiting to maximum point.
Indicates that the sweep/list manual point has been reassigned
to a smaller number value due to the longest list decreasing in
size or being turned off. Its new value is the length of the
frequency list which is the longest enabled list.
Data out of range;Manual point exceeds power
list size. Limiting to maximum point.
Indicates that the sweep/list manual point has been reassigned
to a smaller number value due to the longest list shrinking, or
being turned off. Its new value is the length of the power list,
which is the longest enabled list.
-223
Too Much Data;The number of list points exceeds
the maximum allowed.
Indicates that a SCPI list has been entered that is longer than
the maximum allowed length, which is also the maximum
number of step points; too many points were given for a
frequency, amplitude, or dwell time list. This error can also be
caused by attempting to copy items in the list editor when the
list is already at its maximum length.
-230
Data corrupt or stale;RAM copy of <filename>.
The non-volatile RAM copy of a file is either corrupt or is out
of date with the EEPROM master copy (if one exists). The
system automatically re-initializes the file from EEPROM (if
appropriate) or from a default algorithm. A potential cause is a
failing backup battery.