3-40
Troubleshooting
ESG Series Signal Generator Instrument-Specific Error Messages
-315 Configuration memory lost;Persistent state
preset. Using factory defaults.
Indicates that the persistent state has been forced to return to
factory preset values.
Configuration memory lost;Persistent state
version is bad. Using factory defaults.
Indicates that the persistent state version is not recognized as
valid and is assumed to be corrupt. The persistent state is
reinitialized with the factory preset values.
Configuration memory lost;Persistent state
checksum is bad. Using factory defaults.
Indicates that the persistent state is corrupt and had to be
reinitialized with the factory preset values.
-321
Out of memory;Unable to verify instrument state
file.
Indicates that an instrument state file could not be accessed and
verified because of insufficient memory. Reduce the size of any
sweep lists and try again.
Out of memory;Memory catalog failed.
Indicates that there is not enough memory to complete a catalog
listing. Reduce the size of any sweep lists and try again.
Out of memory;Unable to display timeslot
window.
Indicates that the instrument was unable to create part of the
graphical user interface due to an inability to allocate memory
(possibly due to fragmentation). Please report the
circumstances to the factory. The instrument is still functional.
Out of memory;Unable to display protocol
window.
Indicates that the instrument was unable to create part of the
graphical user interface due to an inability to allocate memory
(possibly due to fragmentation). Please report the
circumstances to the factory. The instrument is still functional.