3-39
Troubleshooting
ESG Series Signal Generator Instrument-Specific Error Messages
File name error;Only one “:” is allowed.
Indicates that only one colon is allowed in any filename
specification. The text before the colon is a user memory
subsystem. The valid user choices are
:, DEFAULT:, STATE:,
and
LIST:.
File name error;Only one “@” is allowed.
Indicates that only one @ is allowed in any filename
specification. It specifies the memory subsystem that a user file
actually resides in.
-286
Program runtime error;Floating-Point Exception
Indicates that a floating-point math error (such as a divide by
zero) has been detected. The system will attempt to recover
automatically. Report the circumstances to the nearest HP Sales
and Service office.
-310
System error;RS232 buffer overflow: character
lost.
Indicates that the RS232 buffer has been exceeded. The most
recent character has been dropped.
System error;Cannot change manual point until
list mode error condition cleared.
An error is keeping the sweep/list from being able to set the
frequency and/or power. Until the problem is addressed, the
manual point cannot be changed.
System error;Unable to determine which
attenuator is installed.
Indicates that an invalid attenuator identification code has been
detected. Possible causes include a loose attenuator control
cable. The instrument will likely not produce the proper output
power levels. Report this error to the factory.
-311
Memory error;Unable to configure Save Recall
registers from non-volatile memory. Save Recall
registers re-initialized.
Indicates that saved states are no longer usable. Delete
explicitly using Catalog.