3-42
Troubleshooting
ESG Series Signal Generator Instrument-Specific Error Messages
Out of memory;Unable to check Data Generator
memory.
There was not enough memory in the system to properly
complete the data generator memory test. This does NOT imply
a data generator memory failure. Check all other error
messages to identify possible causes, discontinue list/sweep
mode to free some memory, and repeat the test.
-330
Self-test failed;Power supply self-test
failure
Indicates that the self-test for a particular power supply voltage
has failed. The instrument is likely not functional. Contact the
nearest HP Sales and Service office.
Self-test failed;EEPROM header checksum error
<card_name>.
Indicates that the card identification header for a hardware card
is incorrect. If the card is not properly identified, the instrument
is likely to be non-functional. Contact the nearest HP Sales and
Service office.
Self-test failed;Data Generator Memory Test @
0x____
Indicates that the data generator memory failed. Modulation
data produced by the data generator may not be correct.
However, if an
Unable to check Data Generator Memory
error was also seen, this result is not conclusive. The address of
the first location that failed is reported. Contact the nearest HP
Sales and Service office.
Self-test failed;Burst Generator Memory Test @
0x____
Indicates that the burst generator memory failed. Modulation
data produced by the burst generator may not be correct.
However, if an
Unable to check Burst Generator Memory
error was also seen, this result is not conclusive. The address of
the first location that failed is reported. Contact the nearest HP
Sales and Service office.