3-43
Troubleshooting
ESG Series Signal Generator Instrument-Specific Error Messages
Self-test failed;Bad address position @ 0x____
Indicates that the data generator memory failed. Modulation
data produced by the data generator may not be correct. An
address that appeared to have a failed address line was reported.
Contact the nearest HP Sales and Service office.
Self-test failed;Chips ___, ___ aliased @
0x____
Indicates that the data generator memory failed. Modulation
data produced by the data generator may not be correct. An
address that appeared to be aliased across multiple memory
chips has been reported. Contact the nearest HP Sales and
Service office.
-430
Query DEADLOCKED
Indicates that a SCPI output queue has filled preventing further
SCPI command execution, and there is no more room left in the
corresponding SCPI input queue to accept a query to read from
the output queue. The system automatically discards output to
correct the deadlock.
201
Bad file number;Unable to check Data Generator
memory.
Indicates that the instrument was not able to generate the
pattern necessary to perform the data generator memory test.
This does NOT imply a data generator memory failure. Report
the problem to the factory.
208
I/O error;Unable to delete saved state from
non-volatile memory. No instrument state
change.
Indicates that a STATe: file could not be deleted due to the file
not being found, file corruption, or another file-related
problem. If the file is displayed by a memory catalog, delete it
explicitly.
I/O error;Save a state register ignored.
Indicates that a STATe: file could not be saved due to
insufficient space, file corruption, or another related problem.