10 - 58
HAPSITE ER Operating Manual
The default window that will be displayed is the Internal Standard Calibration
Response Table. See Figure 10-76. This table will display the file names, the area
of the peak and the response factor (the ratio between the signal produced by the
analyte and the quantity of the analyte which produces a signal).
Figure 10-76 Internal Standard Calibration Response Table
By selecting analyte from the drop-down menu, the Analyte Standard Calibration
Response Table will be displayed. The Analyte Standard Calibration Response
Table contains the same information as the Internal Standard Calibration Table,
but pertains to the calibrated analytes. See Figure 10-77.
Figure 10-77 Analyte Standard Calibration Response Table