11 - 7
HAPSITE ER Operating Manual
18 The highlighted compound in the library template will correspond to the peak
with the red dotted line. (See Figure 11-18.)
Figure 11-18 Peak Correspondence
19 Verify that all analytes have a Net Fit greater than 70 by hovering the mouse
over the analyte name. (See Figure 11-19.)
Figure 11-19 Net Fit
20 Verify that the retention times are correct. (See Figure 11-20.)
Figure 11-20 Retention Time Verification