3 - 28
HAPSITE ER Operating Manual
18 Using a 7/16 in. wrench, turn the top nut approximately 1/4 of a turn counter 
clockwise until tight. (See Figure 3-51.)
Figure 3-51  Using Wrench to Tighten Top Nut
19 When gentle upward pressure is applied to the top elbow, the elbow should not 
move. If the elbow moves, the concentrator is not properly seated. Repeat 
section 3.3.7, Replacing the Concentrator, on page 3-21.
20 The concentrator cover contains two metal contacts. Inspect the contacts to 
verify that they fit snuggly onto the concentrator collars, ensuring electrical 
contact, without placing uneven strain on the glass concentrator tube. (See 
Figure 3-52.)
Figure 3-52  Inspecting Concentrator Cover