4 - 23
HAPSITE ER Operating Manual
14 To confirm the Survey results with a GC/MS run, ANALYZE can be touched or
ANALYZE RUN can be pushed during the Survey run. (See Figure 4-41.)
NOTE: It is advised to begin an Analyze run either after a peak has been
displayed and/or Survey has been run for the full two minutes.
15 If running an Analyze method is not desired, touch STOP to stop the sampling
process and automatically save the data. (See Figure 4-41.)
Figure 4-41 ANALYZE and STOP
16 A METHOD FINISHED message will appear when the Survey method has
ended. (See Figure 4-42.)
NOTE: The total time required for a Survey analysis is typically less than 3
minutes. If the Survey is not stopped manually, it will automatically stop
at 5 minutes.
Figure 4-42 Survey Method Finished