EasyManua.ls Logo

Inficon HAPSITE ER - Page 82

Inficon HAPSITE ER
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 - 22
HAPSITE ER Operating Manual
12 To view the chromatogram while the method is running, touch GRAPH. (See
Figure 4-39.) Alternately, use the arrow keys to highlight GRAPH and push
OK SEL.
NOTE: This screen will also state the TIC Max, the current TIC and the time
the method has been running.
NOTE: Touching the blue compound identification above the chromatogram
will display its Synonym and Exposure Limit information.
Figure 4-39 Sample GRAPH function
13 When the TIC begins to increase, move the probe away from the sample of
interest. Continue the run until the TIC level returns to the initial background
TIC level that was noted in Step 8 on page 4-20. (See Figure 4-40.)
NOTE: Monitor the side bar on the HAPSITE ER screen for guidance. The bar
rises as the TIC increases and green signifies that the proper sampling
distance is being maintained. To avoid saturation, remove the probe
from the sample when the bar increases and turns yellow. If saturation
occurs, the side bar will turn red and the TIC will be above 60 million.
Figure 4-40 Returning to Baseline

Table of Contents

Related product manuals