Page #37 background image
Keysight B2961A/B2962A SCPI Command Reference, Edition 6 1- 19
Programming Basics
Status System Diagram
Figure 1-3 Questionable Status register
Notes:
C = Condition register
T = Transition filter
EV = Event register
EN = Enable register
C
EV EN
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
< 1 >
< 2 >
< 4 >
< 8 >
< 16 >
< 32 >
< 64 >
< 128 >
< 256 >
< 512 >
< 1024 >
< 2048 >
< 4096 >
< 8192 >
< 16384 >
Not Used
Questionable Status
:STAT :QUES :COND ?
:STAT :QUES :EVEN ?
:STAT :QUES :ENAB?
:STAT :QUES :ENAB
Voltage Summary
Self-Test Summary
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
:STAT :QUES :NTR <enable>
:STAT :QUES :NTR ?
:STAT :QUES :PTR <enable>
:STAT :QUES :PTR ?
T
Current Summary
Temp. Summary
Interlock
0
1
2
3
…
13
14
15
Channel 1
Channel 2
:STAT
:QUES :VOLT ?
:STAT:QUES :CURR ?
(Same as Voltage Summary )
:STAT :QUES :TEMP ?
(Same as Voltage Summary )
Calibration Summary
Same as Voltage Summary )
Same as Voltage Summary )
:STAT :QUES :CAL ?
:STAT :QUES :TEST ?
(
(
Ch1 Output Protection
Ch2 Output Protection
Ch1 Transient Event Lost
Ch2 Transient Event Lost
Ch1 Acquire Event Lost
Ch2 Acquire Event Lost

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals