Keysight B2961A/B2962A SCPI Command Reference, Edition 6 2- 7
Subsystem Command Summary
Setting Source Output and Measurement
[:SOUR[c]]:ARB:CURR:EXP:TCON
time
[:SOUR[c]]:ARB:VOLT:EXP:TCON
time
[:SOUR[c]]:ARB:CURR:EXP:TCON?
[DEFault|MINimum|MAXimum]
[:SOUR[c]]:ARB:VOLT:EXP:TCON?
[DEFault|MINimum|MAXimum]
Set the time constant of the exponential
waveform output.
time=MINimum|MAXimum|DEFault|0 to
1000 seconds.
0
[:SOUR[c]]:ARB:CURR:EXP:TIME
time
[:SOUR[c]]:ARB:VOLT:EXP:TIME
time
[:SOUR[c]]:ARB:CURR:EXP:TIME?
[DEFault|MINimum|MAXimum]
[:SOUR[c]]:ARB:VOLT:EXP:TIME?
[DEFault|MINimum|MAXimum]
Set the time of the exponential waveform
output. Waveform period must not exceed
1000 seconds.
time=MINimum|MAXimum|DEFault|0 to
1000 seconds.
0
[:SOUR[c]]:ARB:CURR:RAMP:END[:LE
V] level
[:SOUR[c]]:ARB:VOLT:RAMP:END[:LE
V] level
[:SOUR[c]]:ARB:CURR:RAMP:END[:LE
V]?
[DEFault|MINimum|MAXimum]
[:SOUR[c]]:ARB:VOLT:RAMP:END[:LE
V]?
[DEFault|MINimum|MAXimum]
Set the end level of the ramp waveform
output.
level=MINimum|MAXimum|DEFault|
minimum to maximum source value, in A
or V. See
“Source Output Ranges” on
page 2-36.
0
[:SOUR[c]]:ARB:CURR:RAMP:END:TIM
E time
[:SOUR[c]]:ARB:VOLT:RAMP:END:TIM
E time
[:SOUR[c]]:ARB:CURR:RAMP:END:TIM
E?
[DEFault|MINimum|MAXimum]
[:SOUR[c]]:ARB:VOLT:RAMP:END:TIM
E?
[DEFault|MINimum|MAXimum]
Set the end time of the ramp waveform
output. Waveform period must not exceed
1000 seconds.
time=MINimum|MAXimum|DEFault|0 to
1000 seconds.
0
Command Summary Reset setting