Keysight B2961A/B2962A SCPI Command Reference, Edition 6 2- 29
Subsystem Command Summary
Setting Source Output and Measurement
:SENS[c]:CURR[:DC]:NPLC:AUTO mod
e
:SENS[c]:VOLT[:DC]:NPLC:AUTO mod
e
:SENS[c]:CURR[:DC]:NPLC:AUTO?
:SENS[c]:VOLT[:DC]:NPLC:AUTO?
Enables or disables the automatic NPLC
function.
mode=1|ON|0|OFF
The automatic NPLC on/off works with
the automatic aperture on/off set by the
:SENS:<CURR|VOLT>:APER:AUTO
command. So the last command setting is
effective for both functions.
OFF
:SENS[c]:CURR[:DC]:PROT[:LEV][:B
OTH]
comp
:SENS[c]:VOLT[:DC]:PROT[:LEV] co
mp
:SENS[c]:CURR[:DC]:PROT[:LEV]?
[DEFault|MINimum|MAXimum]
:SENS[c]:VOLT[:DC]:PROT[:LEV]?
[DEFault|MINimum|MAXimum]
Sets the compliance value of the specified
channel. The setting value is applied to
both positive and negative sides.
comp=MINimum|MAXimum|DEFault|
minimum to maximum measurement
value, in A or V. See maximum current or
maximum voltage in
“Source Output
Ranges” on page 2-36.
+1.00E-04
A for current
compliance,
+2.0 V for
voltage
compliance
:SENS[c]:CURR[:DC]:PROT[:LEV]:NE
G
comp
:SENS[c]:VOLT[:DC]:PROT[:LEV]:NE
G
comp
:SENS[c]:CURR[:DC]:PROT[:LEV]:NE
G?
[DEFault|MINimum|MAXimum]
:SENS[c]:VOLT[:DC]:PROT[:LEV]:NE
G?
[DEFault|MINimum|MAXimum]
Sets the negative side compliance value
of the specified channel.
comp=MINimum|MAXimum|DEFault|
minimum to maximum measurement
value, in A or V. See maximum current or
maximum voltage in
“Source Output
Ranges” on page 2-36.
-1.00E-04 A
for current
compliance,
-2.0 V for
voltage
compliance
:SENS[c]:CURR[:DC]:PROT[:LEV]:PO
S
comp
:SENS[c]:VOLT[:DC]:PROT[:LEV]:PO
S
comp
:SENS[c]:CURR[:DC]:PROT[:LEV]:PO
S?
[DEFault|MINimum|MAXimum]
:SENS[c]:VOLT[:DC]:PROT[:LEV]:PO
S?
[DEFault|MINimum|MAXimum]
Sets the positive side compliance value of
the specified channel.
comp=MINimum|MAXimum|DEFault|
minimum to maximum measurement
value, in A or V. See maximum current or
maximum voltage in
“Source Output
Ranges” on page 2-36.
+1.00E-04
A for current
compliance,
+2.0 V for
voltage
compliance